You are here: Home > X-ray>Fixed x-ray>High Frequency X-ray machine

High Frequency X-ray machine

BT-XR07 High Frequency X-ray Radiography machine



1. Application:

This machine is applied to take radiography on every part of human body, such as head, limbs, chest, limbus and abdomen and etc.




2. Specifications

Item

content

Output power

50 kW

Main inverter frequency

30 kHz

X-ray tube

Rotary anode, dual-focus

Large/ Small focus

0.6mm/ 1.2mm

Tube thermal capacity

212kJ

Tube current

32-630mA Step adjustment

Tube voltage

40kV-150kV 41 Steps adjustment

Time

0.003-6.0 s

Table radiography grids

(Imported)

Grid density: 103L/inch

Grid ratio: 10:1;

Focusing Distance: 100cm

Stationary type: 15"×18"

Mechanical Movement

SID:

500-1280mm

X-ray tube Longitudinal movement

1300mm

X-ray tube Rotation range around transverse axis

±90°

X-ray tube Rotation range around pillar

4* 90

Table

Table size

2000×760mm

Table height

≤700mm

Table transverse movement

220mm

Table longitudinal movement

900mm

Cassette Movement Distance

500mm

Light Time of limiter

15s

Film size

57"-14"×17"

Power supply

380V three phase 50Hz

75kVA

3.Standard Configuration

No.

Item

Quantity

1

Control console     New integral

1

2

High frequency and high voltage X-ray generator

1

3

X-ray tube

1

4

High voltage cable

1

5

Radiography table (with pillar)

1

6

Manual limiter

1

7

BT-PLD5800B accessory

1


Option: Bucky stand

(Grid density: 103L/inch, Grid ratio: 10:1, Focusing Distance: 180cm,

Stationary type: 15"×18", film size: 14"×17", Up and down movement: 480-1460mm)



IV. Feature

The console applies LCD display photograph term

The console control the expose time by MCU internal clock, make the expose time accurate

Adopt microcomputer monitoring start-up and operation

Adopt mA, KV, s three parameters free matching work method

The low-voltage power supply was isolated in this machine

Hand-switch control module

External interlock device interface

This machine has temperature control switch to protect the x-ray tube